Calibration and validation strategy for electromechanical cardiac digital twins

  1. Department of Computer Science, University of Oxford, Oxford, United Kingdom
  2. Barcelona Supercomputing Centre, Barcelona, Spain
  3. ELEM Biotech, Barcelona, Spain

Peer review process

Revised: This Reviewed Preprint has been revised by the authors in response to the previous round of peer review; the eLife assessment and the public reviews have been updated where necessary by the editors and peer reviewers.

Read more about eLife’s peer review process.

Editors

  • Reviewing Editor
    Patrick Boyle
    University of Washington, Seattle, United States of America
  • Senior Editor
    Olujimi Ajijola
    University of California, Los Angeles, Los Angeles, United States of America

Reviewer #1 (Public review):

Summary:

The study by Wang et al. investigates cardiac electromechanical modeling and simulation techniques, focusing on the calibration and validation of ventricular models according to ASME V&V40 standards. The researchers aim to calibrate model parameters to align with key biomarkers such as QRS duration and left ventricular ejection fraction and validate the model against independent measurements such as displacement and strain metrics. The authors also examine the impact of parameter variations on deformation, ejection fraction, strains and other biomarkers. The overarching aim of the study is to give credibility to the underlying computational electromechanics framework as a step towards the cardiac Digital Twin vision.

Strengths:

(1) The study presents a solid validation strategy for cardiac models based on independent data.

(2) It integrates electrophysiological, mechanical, and hemodynamic biomarkers for sensitivity analysis and calibration.

Weaknesses and Limitations:

(1) Model Assumptions: The study relies on several simplified modeling assumptions that do not reflect the current state-of-the-art:

a) Isotropic scaling of the ventricular mesh to generate an unloaded reference geometry.

b) Simplified afterload and preload models that do not consistently capture the full range of physiological responses.

c) Simplified epicardial boundary conditions.

These limitations are appropriately acknowledged and discussed by the authors in a dedicated Limitations section.

(2) Numerical Framework:

a) The numerical framework used for the mechanical part of the model may be susceptible to locking effects that could contribute to artificially stiff and less contractile behavior. This is indicated by a ten-fold scaling of the peak active contractile force parameter relative to literature values and notable sensitivity of the model to the tissue compressibility parameter. While - as acknowledged by the authors - part of this can be attributed to simplified modeling choices, other comparable studies have not reported similar issues.

b) The human electrophysiology model is not described in enough detail. Currently, it is not mentioned in the manuscript that an Eikonal model was used to compute activation times on the endocardial surface to be robust against coarse mesh resolutions.

(3) Geometrical model and digital twin: The model presented combines anatomical data, electrical measurements, and physiological reference values from different individuals or population averages, rather than being derived from a single patient. The authors have appropriately moderated their claims in the revision, framing the work as a step towards the cardiac Digital Twin vision rather than asserting that the model itself constitutes a digital twin.

(4) Calibration procedure: The description of the calibration procedure has been substantially improved in the revision. The authors now provide explicit rationale for each calibration step and clarify that the procedure targets multiple physiological biomarkers in sequence. Verification that the calibrated model produces physiological cellular dynamics, including intracellular calcium transients, is now provided. The revised manuscript also shows the simulated electrocardiogram alongside population reference ranges, which partially addresses the question of calibration quality. However, a direct comparison of the simulated electrocardiogram with the individual measured signal used for calibration is not provided, which would give a more stringent and direct assessment of how well that specific calibration target was achieved.

Comments on revised version.

The revision represents a genuine improvement. The calibration procedure is now more transparently described, physiological cellular dynamics are verified, the digital twin framing has been appropriately moderated to reflect a step towards that vision rather than a claim of having achieved it, and an expanded limitations section identifies where the framework falls short.

Several of the concerns raised in the first round have been addressed, but some issues remain:
The model still requires a ten-fold scaling of the peak active contractile force relative to literature values, and the imbalance between left and right ventricular output persists, along with non-physiological right ventricular pressures and ejection fraction. These are partly fundamental limitations of the current modelling approach that may not be fully resolvable within the scope of this paper, and the authors are to be credited for acknowledging them. However, they do constrain the conclusions that can be drawn about the credibility of the framework for reproducing healthy cardiac physiology.

The population-averaged reference dataset and the calibration and validation framework remain contributions of value to the community, and the revised limitations section adds useful transparency about the current state of the art.

Reviewer #2 (Public review):

The authors present an interesting study on calibrating and validating a biventricular cardiac electromechanical model. This is an important contribution, but some questions remain about the quantitative validation and verification aspects of the study.

Major comments:

(1) The title and paper stress the importance of validation on several occasions. However, the actual validation performed is limited to the section in lines 427-439. Furthermore, it is entirely qualitative, making assessing the model's quality difficult. Most of the paper is focused on sensitivity analysis, which is also interesting but unrelated to validation. Can you include a quantitative comparison with deformation biomarkers? E.g., spatially quantify strain differences between simulation and in vivo data, or overlay the current configuration of the geometry with MRI in various views, and calculate a displacement error norm.

(2) You mention the ASME V&V40 standards throughout your paper. Yet, you only address the "second V" validation, ignoring the "first V" verification. How did you ensure that your computational models are implemented correctly?

(3) All parameters discussed in this publication are physical parameters. What is the sensitivity of your model outputs concerning computational parameters?

Comments on revised version.

The authors have addressed my prior comments

Author response:

The following is the authors’ response to the original reviews.

We greatly appreciate the reviewers for their efforts in reviewing our manuscript. We highlight that the key contributions of our paper are to provide a framework for calibration and validation of high-fidelity cardiac electromechanical models based on a diverse compilation of clinical datasets, and that we provide one example of such an evaluation of our own baseline electromechanical model. The comments raised by the reviewers were chiefly focused on the second goal, which is our specific model and the outcomes of the evaluation process, rather than on the evaluation framework itself. As such, we have made improvements to our model implementation and to provide additional confidence in our specific modelling framework through this review process. Specifically, we have strengthened the verification component of this evaluation, provided additional quantitative measures, and included a more in-depth discussion of the remaining limitations in our modelling framework. We hope that the updated version of the manuscript and our efforts to improve it are well-received by our reviewers and editors, as well as by the modelling and simulation community at large.

eLife Assessment

This is a potentially important study that explores the relevant range of parameter values for calibration and validation of cardiac electromechanics in ventricular models. Although much of the work presented is solid, the evidence provided to support the authors' key scientific claims is incomplete, especially as it relates to the emphasis on standardized validation and verification approaches. Notably, the level of model personalization presented in this work falls short of the threshold for what could reasonably be called a "digital twin", even by the relatively relaxed standards that have emerged in computational physiology and related fields in recent years.

We appreciate the eLife assessment for identifying the potential importance of our study. Regarding the threshold for 'digital twin', we note that a cardiac digital twin is envisioned as a patient-specific computational model of the heart, personalised from multi-modal clinical data and continuously updated to support diagnosis, prognosis, and treatment planning, which is a goal that to our knowledge no published electromechanical study has simultaneously fulfilled. It is for this reason that the community refers to the 'digital twin vision' rather than its realisation, and we adopt this framing consistently throughout the manuscript.

The primary contribution of this manuscript is the framework: a systematic application of ASME V&V40 standards to a fully coupled electromechanical model, spanning electrical, mechanical, and haemodynamic biomarkers within a single study. In our revision, we have clarified that the model evaluation presented here is an example application of that framework, which was designed not to certify a model as complete, but to provide a transparent audit of current capability that identifies where confidence is established and where further development is needed. We have updated the title and language throughout the manuscript to reflect this framing consistently.

Public Reviews:

Reviewer #1 (Public review):

Summary:

The study by Wang et al. investigates cardiac electromechanical modeling and simulation techniques, focusing on the calibration and validation of ventricular models according to ASME V&V40 standards. The researchers aim to calibrate model parameters to align with key biomarkers such as QRS duration and left ventricular ejection fraction, and validate the model against independent measurements such as displacement and strain metrics. The authors also examine the impact of parameter variations on deformation, ejection fraction, strains, and other biomarkers. The overarching aim of the study is to give "credibility to the underlying computational electromechanics framework" and to "pave the way towards credible cardiac electromechanical Digital Twins."

Strengths:

(1) The study presents a solid validation strategy for cardiac models based on independent data.

(2) It integrates electrophysiological, mechanical, and hemodynamic biomarkers for sensitivity analysis and calibration.

Weaknesses and Limitations:

(1) Model Assumptions: The study employs simplified modeling assumptions that are not state-of-the-art, e.g.,

(a) Isotropic scaling of the mesh to generate an unloaded reference geometry.

(b) Simple afterload and preload models that fail to produce physiological results.

(c) Simplified epicardial boundary conditions.

While our model was able to broadly achieve physiological behaviour based on the calibration and validation datasets, it also contains several simplifications that can be expanded with more sophisticated techniques to allow explorations in specific areas. We have added a dedicated Limitations subsection to the Discussion section of the manuscript to address these and to provide references to relevant studies.

(2) Numerical Framework:

(a) The mesh resolution and/or the numerical framework used for the mechanical part appears to suffer from known numerical artifacts (locking effects), leading to overly stiff or inaccurate behavior in finite element analysis. This results in an artificially stiff response to deformation, which is compensated by setting active contraction to ten times the value reported in the literature. The authors attribute this to limitations in using ex vivo tissue measurements to represent in vivo function, although similar issues were not observed in previous works.

We thank the reviewer for raising this point and have investigated it carefully. We have added a verification section as well as discussions to the manuscript to more comprehensively discuss this point. In short, through various tests against benchmark (Land) and comparing stress-strain curves in cube simulations with the same mesh resolution, we could not identify evidence of volumetric locking effects. The elevation in contractile force was also necessary in a simplified ellipsoid version of the model in a previous publication [ref 7, Levrero-Florencio, et al. 2020]. We note that these benchmarks were performed in the incompressible transversely isotropic regime; whether analogous locking effects exist in the dynamic orthotropic active contraction framework used in the full biventricular simulations remains an open question, which we have identified as a priority for future benchmarking, for example against the Arostica et al. 2025 benchmark.

We agree that the explanation of this as ex vivo vs in vivo difference in contractile force is too simple, and other contributing factors are better understood through comparison with similar studies in the field. Strocchi et al. (2023) used a four-chamber model with explicit atrial mechanics, and in her history matching varied Tref within +- 33-55% of a reference value of 120-150 kPa, targeting a peak active tension of 160 +- 15 kPa, which was a considerably more modest adjustment than applied here, likely reflecting differences in model geometry, pericardial constraint, and circulatory model between the two studies. Gerach et al. (2021, Mathematics) applied manual parameter adjustments informed by in vivo active tension measurements of 120 – 150 kPa and achieved ejection fractions of approximately 63%; however, they reported that systolic pressures in both ventricles were too high for a healthy heart, and similarly reported elevated peak ejection rates compared to MRI measurements, a difficulty we also encountered. Notably, Gerach et al., report that atrial contraction contributes approximately 11-13% of end-diastolic volume, which in a biventricular-only model would directly reduce the achievable LVEF and necessitate compensating adjustments to active tension. Zingaro et al. (2024, Journal of Computational Physics), using an alternative active tension model (RDQ20), similarly found it necessary to increase contractility parameter (a_XB) to achieve sufficient ejection, and explicitly report that no single parameter configuration simultaneously achieved physiological peak ejection rate and LVEF, a fundamental tension we also encountered. Together, these comparisons suggest that the elevated Tref in our model most likely reflects a combination of the absence of atrial filling, simplifications in pericardial constraint, and the lack of poroelastic behaviour, rather than volumetric locking alone. Additional investigations are needed as explained in the manuscript, and these factors are identified as open priorities for future development within our framework.

We added a comparison with these three studies to the discussion section of the manuscript and have updated the limitations section to reflect this more nuanced account of the factors contributing to the elevated active tension scaling. Further work will be required to address these points.

(b) Further, the authors employ the monodomain model for the simulation of the electrical excitation and relaxation on a relatively coarse grid with an approximate edge length of 1mm. This resolution is known to be insufficient for reliable results in organ-scale electrophysiology modeling.

Our ECG simulations are robust against coarse mesh resolutions since we use an Eikonal solution to prescribe the activation times on the endocardial surface and we tune the diffusivity parameters in the model such that the correct conduction velocities are reached, as performed in Camps et al. (2024) (ref 33) using the tuneCV tool in monoAlg3D (https://github.com/rsachetto/MonoAlg3D_C/tree/master/scripts/tuneCV), which is similar to the tool in openCARP (described here: https://opencarp.org/documentation/examples/02_ep_tissue/03a_study_prep_tunecv).

While this mesh resolution may not be sufficient for simulations of more complex behaviour, such as re-entry and fibrillation patterns, it is sufficient for simulations of ECGs in this study. We have noted this in the methods section.

(3) Geometrical model and digital twin: The geometrical model, taken from a public cohort and calibrated to an ECG of another individual along with population-averaged values from a databank (UK Biobank), and unrelated measurements from surgical procedures, can hardly be considered a digital twin. Further, validation of the model was then performed against data from yet another cohort.

We thank the reviewer for this point and welcome the opportunity to clarify our dataset choices. The use of multiple data sources was a deliberate methodological decision. While an ideal dataset for electromechanical model evaluation would combine full biventricular geometry, 12-lead ECG, invasive pressure measurements, and myocardial strain data from a single individual, no such dataset currently exists in the public domain, and acquiring it routinely would be impractical in clinical settings. Multi-source integration therefore reflects the realistic deployment scenario for future clinical translation of these tools.

The specific choice of geometry was principled: the mesh associated with the ECG dataset that was available to us was truncated at the base due to the clinical acquisition protocol, which would have prevented physiologically realistic basal boundary conditions. The female Rodero geometry we chose provides full ventricular coverage and was selected on that basis.

Demonstrating that a coherent, systematically evaluated framework can be constructed from compiled multi-modal data is itself a contribution because it makes the tools accessible to the wider community without requiring a single ideally acquired dataset.

(4) Calibration procedure: There are apparent flaws in the calibration procedure, or it is not described in sufficient detail. The authors dedicate significant effort to motivating parameter ranges, but in the end they use mostly other parameters for the calibration process, aiming to maximize left ventricular ejection fraction. It is not clear whether the chosen parameters result in, e.g., physiological calcium traces or calibrated parameters that are within physiological ranges.

Thank you for raising this point, which we have now clarified in the manuscript. The parameters that were chosen for the calibration process were based on the results of the sensitivity analyses.

In addition, we have supplemented results Figure 1 with a subfigure F showing that the calcium transient and action potential durations fall within physiological ranges after calibration.

(5) Goodness of fits, e.g., a direct comparison of the measured and the simulated ECG, are not provided to assess calibration quality.

The calibrated model achieves QRS duration of 89 ms and QT interval of 360 ms, both of which fall within the healthy reference ranges compiled in Table 2, providing a biomarker-level assessment of calibration quality (Figure 1A). A full quantitative goodness of fit analysis of the simulated ECG morphology was performed following the methodology of Camps et al. [52], in which the same beat-averaged ECG was processed; we direct the reader to that work for full details rather than reproducing the analysis here.

(6) Due to these limitations and weaknesses, the authors fall short of achieving some of their goals, particularly establishing credibility for the underlying computational framework and in reproducing healthy pressure-volume loops, and in achieving physiological simulations while using physiological or reported ranges for the calibrated parameters.

For example, a key physiological requirement is that the right and left ventricular stroke volumes are approximately equal in a heart beating at a limit cycle, as the blood pumped by the right ventricle into the pulmonary circulation must match the amount pumped by the left ventricle into the systemic circulation. This balance is not achieved in this study.

We thank the reviewer for identifying the stroke volume imbalance. We acknowledge the physiological requirement that, in a steady-state limit cycle, the right ventricular stroke volume must approximately equal the left ventricular stroke volume. However, since our model does not explicitly prescribe volumes, to achieve this, we would need to either explicitly tune active tension for the left and right ventricles separately, such as done in https://www.frontiersin.org/journals/physiology/articles/10.3389/fphys.2021.716597/full or develop a more sophisticated circulatory model and employ a multistep procedure that sequentially tunes circulatory dynamics, passive mechanics, and active contraction, such as done in https://www.biorxiv.org/content/10.64898/2025.12.11.693778v1.full. Both of which are beyond the scope of this paper.

We note that, despite the absence of explicit RV calibration, the RV volumetric measures and pressures remain within physiological ranges, suggesting that the coupled biventricular mechanics are broadly plausible. As such, we have noted this limitation in our discussion section, and sign-posted to other studies where the stroke volume match is achieved.

(7) The conclusive claim that "the study paves the way towards credible electromechanical cardiac Digital Twins" is not supported. The model exhibits non-physiological behavior, requires unsupported parameter alterations (such as a 10-fold active stress scaling), and does not represent a digital twin, as model data are drawn from various unrelated, non-patient-specific sources.

We thank the reviewer for this comment, which gives us the opportunity to clarify our use of the term 'digital twin'. A cardiac digital twin is envisioned as a patient-specific computational model of the heart, personalised from multi-modal clinical data and continuously updated to support diagnosis, prognosis, and treatment planning. This is a transformative goal for precision cardiology that the field is actively working towards, with credible, systematically validated electromechanical models as its essential foundation. To our knowledge, no published study in cardiac electromechanical modelling has simultaneously fulfilled all three requirements, and it is for this reason that the community often refers to the 'digital twin vision' rather than its realisation.

The primary contribution of this manuscript is the framework: a systematic application of ASME V&V40 standards to a fully coupled electromechanical model, spanning electrical, mechanical, and haemodynamic biomarkers in a single study. The model evaluation presented here is an example application of that framework. Importantly, the framework is not designed to certify a model as complete, but to provide a transparent audit of current capability by identifying where confidence is established and where further development is needed. In this sense, the limitations surfaced through this evaluation are themselves a contribution: they define open problems and priorities for the field.

We have added a definition of the digital twin concept and the roadmap towards its realisation to the introduction and have updated the language throughout the manuscript to consistently reflect the distinction between the framework contribution and the model evaluation. We maintain that this transparent approach represents a meaningful step towards the digital twin vision.

The specific limitations of the current model implementation are addressed in detail in the relevant sections of this response and in the updated manuscript, where we have substantially strengthened the verification and discussion components.

Conclusion:

Overall, this reviewer considers that the study requires a major revision, including improvements in numerical methods, modeling choices, and checks for physiological behavior. Nevertheless, the provided tables with averaged values from the UK Biobank and the presented validation strategy could be valuable to the research community.

Reviewer #2 (Public review):

The authors present an interesting study on calibrating and validating a biventricular cardiac electromechanical model. This is an important contribution, but some questions remain about the quantitative validation and verification aspects of the study.

Major comments:

(1) The title and paper stress the importance of validation on several occasions. However, the actual validation performed is limited to the section in lines 427-439. Furthermore, it is entirely qualitative, making assessing the model's quality difficult. Most of the paper is focused on sensitivity analysis, which is also interesting but unrelated to validation. Can you include a quantitative comparison with deformation biomarkers? E.g., spatially quantify strain differences between simulation and in vivo data, or overlay the current configuration of the geometry with MRI in various views, and calculate a displacement error norm.

We thank the reviewer for this comment.

We have strengthened the quantitative aspect of the validation by reporting the peak simulated strain values for each component and comparing them against the physiological ranges compiled in Table 2. Specifically, the simulated peak strains were: E_ff ≈ -0.20, E_cc ≈ -0.15, E_rr ≈ +0.15, and E_ll ≈ -0.23. These show broad agreement with the in vivo reference ranges from Moulin et al. (2021), noting that the reference ranges are derived from a cohort of 30 subjects and therefore represent a relatively narrow population sample. Shortening strains (fibre and circumferential) are in good agreement, while radial strain is underestimated. We have noted this as a limitation. We have also indicated that a further validation would include a fully quantitative spatial comparison, such as a displacement error norm or voxel-wise strain difference map. This would require access to the raw image data and patient-specific geometry registration, which is beyond the scope of the current study.

(2) You mention the ASME V&V40 standards throughout your paper. Yet, you only address the "second V" validation, ignoring the "first V" verification. How did you ensure that your computational models are implemented correctly?

Thank you for raising this point. We have now included a section on model verification to the manuscript at where we perform benchmarking simulation using the Land (2015) passive inflation benchmark. We also provide a mesh subdivision analysis of the final calibrated model. Additional verifications and previous sensitivity analyses using the same numerical scheme with idealised ellipsoid geometries are also referenced in the verification section, to provide additionally confidence.

(3) All parameters discussed in this publication are physical parameters. What is the sensitivity of your model outputs concerning computational parameters?

Numerical analyses for the Alya solver used in this study has previously been published in works including Levrero et al (2021), which performed sensitivity analyses in a truncated ellipsoid geometry, and Santiago et al (2018), which demonstrated mesh convergence in a cantilever. We have updated the manuscript to point the reader to these studies.

Recommendations for the authors:

Reviewer #1 (Recommendations for the authors):

Major concerns:

(1) Active stress scaling:

The initial value for T_ref appears to be 120kPa * 10, which would be ten times the literature value fitted to human contraction data. Additionally, Table 2 lists a range of [1200-2400], which is 10 to 20 times the literature value.

This discrepancy suggests that other model parameters, model assumptions, or the numerical scheme may be inadequate. In contrast, similar calibrations using comparable models (ToRORd-Land) in other works, such as Strocchi et al. [29], yielded T_ref values close to the literature value.

We thank the reviewer for this comment. As discussed in our response to the public review comment 3a, the elevated T_ref scaling warrants explanation.

We note that Strocchi et al. use a four-chamber geometry include atrial mechanics and a different pericardial constraint, any of which could contribute to differences in the required T_ref scaling. The elevated scaling in our model likely reflects a combination of factors including the absence of poro-elastic behaviour, simplifications in pericardial constraint, and the lack of atrial mechanics, rather than volumetric locking alone. We have added text to the discussion acknowledging this more explicitly and have flagged planned additional benchmarking of the dynamic orthotropic scheme as future work.

(2) Non-physiological results, see Figure 1:

In a healthy heart, RV stroke volume should approx. match LV stroke volume. This is clearly not the case in Figure 1B, where the RV EF is also notably low at 35%.

Consequently, the study fails to reproduce healthy pressure-volume loops, undermining its claim to create a credible cardiac electromechanical digital twin. Hence, also the "Question of interest" posed in line 206 must be answered with a clear "No".

Matching stroke volumes should be a primary calibration goal.

We thank the reviewer for this comment. We agree that stroke volume balance is an important physiological criterion, and we have added it explicitly to the framework criteria in the updated manuscript, noting that our current model evaluation does not satisfy it. This is precisely the kind of transparent appraisal the V&V40 framework is designed to produce: a systematic accounting of which criteria are met and which require further development. A framework that only gets applied to models that pass all criteria would be selection-biased and less informative to the community.

However, we respectfully disagree that the question of interest must be answered with a clear 'No'. We draw the reviewer's attention to the quantities of interest defined in the paper, which are predominantly left ventricular biomarkers, reflecting the intended scope of the calibration framework. The framework successfully reproduces these defined quantities of interest, and the LV pressure-volume loops, strain, volumes and ejection fraction are all within physiological ranges and well-matched to reference data. These quantities of interest were selected based on their clinical implications in cardiac diseases, as detailed in Table 3.

We agree that stroke volume balance is an important physiological requirement for a fully calibrated biventricular model, and we have strengthened the future work and limitations section accordingly.

(3) Inadequate numerical framework:

(a) Monodomain model: The geometries from Rodero et al. [25] have an average edge length of 1mm. It is known that such a coarse resolution leads to inaccurate EP results. It is not mentioned if the authors refined that geometry to an appropriate resolution or used an Eikonal model to mitigate this issue.

As explained earlier, our ECG simulations are robust against coarse mesh resolutions since we use an Eikonal solution to prescribe the activation times on the endocardial surface, as performed in Camps et al. (2024), and we tune the diffusivity parameters in the model such that the correct conduction velocities are reached. We have added a figure in the appendix of this manuscript to show that by increasing the mesh resolution by one subdivision, we get virtually identical ECG simulations. While this mesh resolution may not be sufficient for simulations of more complex behaviour, such as re-entry and fibrillation patterns, it is sufficient for this study. We have noted this in the methods section.

(b) Material law:

- recent publications show that an unsplit deformation gradient for the anisotropic contribution is beneficial to reduce locking effects, see, e.g., Gueltekin et al. Computational Mechanics 63, no. 3 (2019): 443-53. https://doi.org/10.1007/s00466-018-1602-9.

- K_ct is a penalty parameter to enforce some degree of incompressibility. Results are highly dependent on the grid size and the finite element formulation due to locking effects.

As the authors write: "In our simulations, we saw that the LVEF was strongly sensitive to changes in the incompressibility of the tissue (Kct), such that an increase in compressibility of the myocardial tissue helped to increase LVEF." Which exactly points to the issue of locking effects.

So an option would be to use a finer grid or a more adequate numerical scheme with quadratic finite elements, as eg. in [5] Fedele et al., or [6] Gerach et al,. or stabilized elements as in Karabelas et al. CMAME 394 (2022) https://doi.org/10.1016/j.cma.2022.114887.

Overall, this does not point to "limitations in using ex vivo tissue measurements to represent in vivo function" but to limitations in the numerical setup. In fact, with an adequate numerical scheme, the simulations should be largely insensitive to the choice of this penalty parameter K_ct. See, e.g., Karabelas et al. above, where the authors varied K_ct from 650kPa to infinity (representing an incompressible material), and there is no visible influence on the PV loops.

We investigated this point using the Alya solver, and we found that the mesh resolution did not alter the LVEF, and our benchmark simulations against Land (2015) did not show the existence of the volumetric locking issue that the reviewer refers to. It is possible, however, that such an effect exists in the elastodynamic orthotropic framework but not in the incompressible and transversely isotropic framework that the Land (2015) benchmarks were set up in. Future analyses could focus on performing additional benchmarking against more recent elastodynamic benchmarks, such as presented in Arostica (2025). We have updated the limitations text in our manuscript to reflect this and to cite relevant literature on this issue.

(4) Boundary conditions:

"This was a simplified version of the method [28], which uses an exponential decay formulation at the 'edge' of the pericardial constraint rather than a step function": I don't really see this in the cited work [28] which gives a spatially varying Robin-type boundary condition at the whole epicardium (i.e. regional scaling of normal springs stiffness based on image-derived motion from CT images) and not only at the edge.

This is motivated by the fact that the pericardial tissue is in contact with various organs of different material properties. Not using spatially varying pericardial parameters is a limitation that might lead to non-physiological deformations, see also Pfaller et al. Biomechanics and Modeling in Mechanobiology 18 (2019): 503-29. https://doi.org/10.1007/s10237-018-1098-4.

We thank the reviewer for this point and we have corrected the manuscript accordingly. To clarify: our implementation applies a uniform Robin spring constraint along the majority of the epicardial surface with zero constraint at the base, which is conceptually similar to Strocchi et al. [28]. The key difference is that Strocchi et al. use a smooth gradient transition from uniform constraint to zero constraint near the base, whereas our implementation uses an abrupt step transition. We acknowledge that a smooth spatially varying transition would more accurately represent the frictionless pericardial contact and have noted this as a limitation in the manuscript with reference to Pfaller et al. [41].

Also check:

- line 117: Gamma_valve_epi is introduced but not used. Was there any boundary condition defined on this valve plane?

- the third equation, maybe (0,T] missing.

- line 120: epicardium instead of endocardium.

These errors have been corrected in the updated manuscript. No boundary conditions were applied on the epicardial surface of the valve plugs, the reference to gamma_valve_epi has been removed.

(5) Reference geometry:

The choice to scale the mesh to a lower volume for the unloading procedure seems questionable. This approach does not ensure that the reloaded mesh aligns with the mesh derived from image data. As a result, the geometry used for the simulations is no longer truly patient-specific.

This mismatch is a significant limitation, as there are established methods available to achieve a proper unloaded configuration, as, e.g., in

Marx et al. Journal of Computational Physics 463 (2022): 111266. https://doi.org/10.1016/j.jcp.2022.111266, and

Regazzoni et al. Journal of Computational Physics 457 (2022): 111083. https://doi.org/10.1016/j.jcp.2022.111083.

As our study aimed at creating a framework for calibration and validation in data-scarce scenarios such as it is often the case in the clinical context, using a compilation of multi-modal data from difference sources, rather than a specific method of personalisation, we did not feel it appropriate to invest significant energy to identify a patient-specific resting geometry, but rather felt that it was important for the resting geometry to fall within population values in terms of diastasis volume. We have clarified this issue in the manuscript and softened claims to Digital Twins in this study. The limitation has been addressed in the updated manuscript, and future work could further address this point.

(6) Calibration procedure:

There are apparent flaws in the calibration procedure, or it is not described in sufficient detail.

We thank the reviewer for raising this point and we have substantially revised the calibration description in the manuscript to clarify the rationale behind each step.

(a) Step 1: "Sample..." Why? kws and Cal50 are not the most significant parameters in the sensitivity analysis. Kct is a penalty parameter dependent on the numerical framework as described above; "ejection pressure threshold" was never mentioned, is it "P ejection LV" in Table 2? Aiming just for the highest LVEF might neglect non-physiological responses to parameter changes.

While kws and Cal50 are not the single most significant parameters for LVEF in isolation, they were grouped in Step 1 because they affect both LVEF and peak systolic pressure simultaneously through cross-bridge cycling rate and residual active tension, making it necessary to sample them jointly rather than sequentially. Kct was included because myocardial compressibility affects wall thickening and therefore stroke volume. The ejection pressure threshold is P_ejection_LV in Table 1 and has now been described explicitly in the methods section. Regarding the concern about non-physiological responses: the action potential duration and active tension were monitored throughout calibration and verified to remain within physiological ranges, as now noted in the manuscript.

(b) Step 2: As systolic pressure is directly dependent on arterial resistance for a 2-element Windkessel model, a uniform sampling approach might not be the best choice here.

We acknowledge that uniform sampling may not be the most efficient approach for Step 2. However, since arterial resistance influences not only peak systolic pressure but also stroke volume and therefore LVEF, a more targeted approach focusing solely on pressure matching could compromise the LVEF achieved in previous steps. Uniform sampling allowed us to select the value that best balanced both quantities simultaneously.

(c) Step 3: The authors mention in line 527: "A four-fold increase in GCaL caused an eight-fold increase in cellular active tension peak". An increase in active tension peak results in higher LVEF. So this step is likely to yield the upper boundary of the GCaL interval.

The reviewer is correct that Step 3 tends to yield a high GCaL value. This was intentional — GCaL was used as a last resort to achieve physiological LVEF after Steps 1 and 2, since the model consistently undershot the target. The upper boundary of the sampled GCaL interval corresponds to a two-fold increase, which remains within the physiological variability bounds applied in previous studies. The resulting action potential duration was verified to remain within physiological ranges.

(d) Step 4: Why again k_ws? It is not the most significant parameter in the SA.

kws was resampled in Step 4 not to increase LVEF further, but to specifically target peak ejection rate and dP/dtmax, which were not adequately matched after Step 3. kws is the dominant parameter affecting these ejection dynamics biomarkers in the sensitivity analysis. Resampling at this stage allowed fine-tuning of ejection dynamics while maintaining the LVEF achieved in previous steps.

(e) Step 5: As far as I can tell, the "diastolic volume change parameter" was mentioned the first time here.

The diastolic volume change parameter C_pLAV has now been described in the methods section in the Phase 5 passive filling description, where it appears as the inverse of the penalty term controlling the rate of return to diastasis volume in the left ventricle.

The whole calibration procedure seems to aim for the highest LVEF, and final values of the calibration parameters are not given.

We note that the calibration procedure does not aim solely for the highest LVEF. As described above, the sequential strategy targets multiple quantities of interest in order of clinical importance: LVEF, peak systolic pressure, peak ejection rate, and peak filling rate, with each step designed to improve a specific subset of biomarkers without compromising those already matched. The final calibrated parameter values are reported in Figure 1F of the revised manuscript.

(7) Novel features in this paper are actually scarce. A way more advanced calibration strategy with a whole heart model, emulators, and also the ToRORd-Land model was already presented in the study by Strocchi et al. [29]. The calibration to ECGs was presented by some of the same authors in Camps et al. [15], and the analysis of cellular effects was already published in several studies by the same group and in other publications, e.g., by the groups of Severi et al.

The systematic compilation of credibility criteria spanning ECG morphology, pressure-volume characteristics, strain and displacement represents a novel contribution in itself, providing the field with a reusable evaluation framework. Furthermore, the present study is designed to yield mechanistic insight into how parameters at different scales influence both electrical and mechanical outputs simultaneously. This goal was not tackled in previous publications, which covered individual components, including ECG calibration in Camps et al. [15] and global sensitivity analysis with whole-heart models in Strocchi et al. [29] with no ECG consideration.

Thus, the work by Camps et al. on ECG calibration was purely electrophysiological and did not investigate the influence of mechanical or haemodynamic parameters on ECG morphology in a fully coupled electromechanical framework. While the effect of mechanical parameters on ECG has been explored by others (e.g. Favino, 2016), this has not previously been examined alongside the relative importance of cellular, mechanical and haemodynamic parameters on pressure-volume characteristics within a single coupled framework. While Strocchi et al. present an emulation strategy, they did not address ECG biomarkers. This distinction is now stated explicitly in the introduction, where we position the present study relative to Camps et al. and Strocchi et al.

(8) How could the calcium sensitivity Cal50 have such a drastic effect on diastolic function, i.e., filling and end-diastolic volume? As far as I understand from the description, the simulation starts with Phase 0 (loading), Phase 1 (atrial filling), and then in Phase 2, electrical activation ensues and active contraction develops, see also the section starting in line 165. Based on this description, I would expect the end-diastolic volumes to be identical across all Cal50 values. Or are the PV loops shown actually limit cycles established over simulations with multiple beats? This point wasn't explicitly clarified in the manuscript.

Calcium sensitivity (Cal50) affects not only systolic active tension development but also diastolic residual active tension, i.e. the degree to which the muscle remains partially activated at end diastole. Higher Cal50 values increase this residual tone, effectively stiffening the myocardium during diastolic filling and reducing end-diastolic volume. This mechanism is well established as a contributor to diastolic dysfunction in heart failure [88]. We have clarified this in the manuscript and also clarified that the PV loops shown are single-beat simulations, not limit cycles, with the end-diastolic volume determined by the prescribed filling pressure alongside the passive and residual active stiffness of the myocardium.

Minor concerns:

(9) Line 29: The values provided: LVEF of 51%, EDV of 110 mL, and ESV of 50 mL are inconsistent. If these values are all related to the LV, the calculated LVEF should be approximately 54.55%, not 51%.

The values quoted in the original abstract were rounded approximation, this has been corrected to report EDV=105 mL and ESV=51 mL, which are consistent with the simulated LVEF of 51%.

(10) "Electromechanical cardiac Digital Twins have had broad applicability ..."

Many of the cited works here are not true "Digital Twins" but rather static, non-patient-specific models of cardiac electromechanics. In some cases, the geometry may be derived from patient data, but this alone does not qualify the model as a digital twin.

This sentence in the introduction has been rephrased as ‘Electromechanical cardiac models have had broad applicability...’. Furthermore, as stated earlier, we have removed explicit claims of Digital Twin from the paper while retaining the fact that this study provides a significant step towards rigorous credibility assessment of the high-fidelity electromechanical models that make Digital Twin construction possible.

(11) While in the abstract and the conclusion, the authors mention "uncertainty quantification", it is mostly a sensitivity analysis that was performed in the paper.

We have updated the text to say ‘sensitivity analysis’ where appropriate in the abstract, results, and conclusion, and replaced ‘uncertainty ranges’ with ‘variability ranges’ throughout. However, since the sensitivity analyses were performed over biologically informed ranges derived from population variability in the literature, the results are informative about how uncertainty in model inputs propagates to uncertainty in simulated biomarkers. We have therefore retained the framing of sensitivity analysis as a first step towards uncertainty quantification in the abstract and conclusion, and have added a clarifying sentence to the methods to this effect.

(12) Line 98: As far as I can tell, the conduction velocity assigned to the endocardial surface - intended to mimic the Purkinje fiber network - is never specified. In the section beginning at line 294, only the transmural conduction velocities are reported.

The endocardial conduction velocity has been specified in the methods section: Purkinje-myocardial junctions were modelled using a fast endocardial activation layer with isotropic conduction velocity of 300 cm/s.

(13) Line 198, Table1:

(a) "21/02/2025 11:09:00 AM" on two occasions is maybe not intended

This has been removed.

(b) For easing up comparisons, units should be consistent between the initial value and the literature ranges, e.g., PV control parameters, heart rate.

Units have been made consistent between the initial values and literature ranges throughout Table 1.

(14) Line 232: "... have already been used to calibrate and validation ...".

This has been corrected.

(15) Line 279, Table 2: This table of variability ranges is not entirely clear and could be improved:

Table 2 has been combined with Table 1 such that the variability ranges sit next to the literature values, for ease of comparison.

(a) "21/02/2025 11:09:00 AM" is maybe not intended.

This has been removed.

(b) use of units should be improved; sometimes it's given in the first column, sometimes in the second column (arterial resistance, compliance), then for k_epi it should be either kPa or kPa/cm.

Units have been made consistent and the units for k_epi has been added in Table 1.

(c) units should also be consistent throughout the paper, e.g. in Figure 1 E arterial resistance is Barye.ms/mL while in Table 2 it is mmHg.ms/mL.

Barye has been removed and replaced by corresponding kPa values throughout the manuscript. This was in the original manuscript since the Alya simulation software were in units of cm, s, g, Barye.

(d) it is also not clear how variability ranges were chosen; e.g., for arterial compliance,e literature ranges are 0.2-2.73 while the chosen range is [0.1,0.2].

The previous ranges were chosen to achieve better LVEF. We have now updated the variability ranges to be purely based on literature values and updated the sensitivity analysis results. The ranges are now presented in Table 1 alongside the literature values for ease of comparison.

(e) For Kct, the initial value in Table 1 is 5000kPa, the literature values are between 10 and 3333, and then the variability range is [10,500]? I guess there is a typo in one of these values.

This has been corrected in the new Table 1.

(f) Table 1 and 2 are in parts redundant.

Table 1 and 2 have been combined into a single new Table 1.

(15) Line 290: It should be uvc_l for the longitudinal coordinate.

This has been corrected.

(16) Line 388, Table 3, regarding values for pressure volume from reference [49]:

(a) the number of participants is 800, including males and females; not only females, see also Table 12 https://jcmr-online.biomedcentral.com/articles/10.1186/s12968-017-0327-9/tables/12

(b) why using female values here while having mixed sex for most of the others? Because the model is female?

The reviewer is correct that reference [49] reports values from a mixed-sex cohort of approximately 800 participants. We used the female-specific values from Table 12 of that reference because the biventricular mesh used in this study was derived from a female subject, making sex-matched reference values the most appropriate comparison. This has been clarified in the manuscript.

(17) Figure 5: What is Jup; why did you choose 0.93 x Jup as reference? Also in Figure 4, why did you use 0.93 x GCal as a reference?

J_up refers to the SERCA<sup2+ reuptake current, which has been relabelled as SERCA throughout the manuscript for consistency. The reference value of 0.93× was used because the sensitivity analysis sampled parameters uniformly between 50% and 200% of baseline using a fixed number of samples, and no sample fell exactly at 1.0×. The closest sampled value was 0.93×, which was therefore used as the reference. This has been clarified in the figure caption.

(18) Line 377: The link to the GitHub repository does not work.

This link has now been made publicly available.

(19) Line 397, Table 3: for the sake of completeness, all abbreviations should be included: e.g., SVL, ESP, EDV, ESV are not included.

This has been written out in full in the new Table 2.

(20) Tick marks in many figures are not readable, e.g., Figure 5 and all the Figures in the appendix.

Tick mark sizes and line widths have been increased across Figures 4, 5, and all appendix figures. The figures have been replotted and updated in the revised manuscript.

Reviewer #2 (Recommendations for the authors):

Minor Comments:

(1) The provided GitHub link https://github.com/jennyhelyanwe/Alya_input_setup/ does not work, potentially because the repository is private. It would be nice to see the repository during the review.

This link has now been made publicly available.

(2) Table 3: Can you include the simulation outputs obtained for validation (with an error indication)? This would summarize the validation that's currently spread out over the results section.

A new Table 3 has been added to the manuscript under the validation section, summarising the simulated values for all deformation and strain biomarkers alongside their reference ranges. The calibration and validate datasets are now reported separately in Tables 2 and 3, respectively.

(3) Figure 1: Add axis labels to all plots.

Axis labels have been added to all subplots in Figure 1 in the revised manuscript. Simulated pseudo-ECG amplitudes are normalised and therefore dimensionless.

(4) Figure 2: Simulated and in vivo strains with exactly the same axes (size, range, ticks) and add grid lines to enable a comparison. Add the mean values of each in the other plot.

The revised Figure 2 now includes the median in vivo strain values from Moulin et al. overlaid as a red dashed reference line on the simulation panels, enabling direct visual comparison. The simulated mean could not be overlaid on the in vivo panels as the original Moulin et al. figure data are not publicly available for replotting. Exact axis matching was not applied as this would cause some simulated curves to fall outside the visible range, obscuring the model behaviour.

(5) Figure 3: The thickness (relative importance of the connections) is impossible to see in this plot. Instead of having gray background connections, remove them entirely below a certain threshold. Make the differences in thickness more pronounced or introduce a continuous color scale for the magnitude of the positive or negative correlation. Alternatively, you could rank the parameters from least to most important in each subfigure A-D and/or provide some numeric values.

Figure 3 has been updated. All non-significant connections (|r| < 0.6 or p > 0.05) have been removed entirely, and gray lines have been removed in each subfigure, making the significant relationships clearer. A continuous blue-to-red colour scale has been applied to indicate the direction of correlation (blue: negative, red: positive), with line thickness proportional to the magnitude of the r-value.

(6) Figure 3 and Table 2: Why were material parameters b, bf, bs, and bfs omitted from this study (but included a, af, as, and afs)?

The b parameters (b, bf, bs, bfs) appear in the exponent of the Holzapfel-Ogden constitutive law and are strongly coupled to the a parameters (a, af, as, afs), which carry units of kPa. In practice, the b parameters can only be reliably identified from ex vivo multiaxial stretch experiments, whereas the a parameters can be estimated from clinical imaging data. Since our study focuses on calibration and validation in a clinical data setting, we included only the a parameters in the sensitivity analysis, consistent with previous personalisation studies.

(7) Figure 4: What do the dotted lines represent?

The dotted lines in Figure 4E highlight the increased longitudinal shortening with increasing GCaL, showing the basal plane moving towards the apex while the apical position remains unchanged due to the pericardial constraint. This has been clarified in the figure caption.

(8) Figures 4, 5, A2-45: Can you use a continuous color scale (e.g., from blue to red) for low to high parameter uncertainty?

A continuous blue-to-red colour scale has been applied to Figures 4, 5, and all appendix figures A2–A6, where blue indicates the lowest parameter value and red indicates the highest. A colour bar has been added to each figure for reference.

  1. Howard Hughes Medical Institute
  2. Wellcome Trust
  3. Max-Planck-Gesellschaft
  4. Knut and Alice Wallenberg Foundation