An In-Silico Testbed for Fast and Accurate MR Labeling of Orthopaedic Implants

  1. Electrical & Computer Eng. Dept. Worcester Polytechnic Institute, Worcester MA 01609-2280 USA
  2. Ansys, Inc., 2600 Ansys Drive, Canonsburg, PA 15317 USA
  3. Dassault Systèmes Deutschland GmbH, Bad Nauheimer Str. 19, 64289 Darmstadt, Germany
  4. Neva Electromagnetics, LLC, 1010 Main St., Holden, MA 01520
  5. GE HealthCare, 500 W Monroe Street, Chicago, IL, 60661 USA
  6. Micro Systems Engineering, Inc., an affiliate of Biotronik, 6024 Jean Road, Lake Oswego, OR 97035 USA
  7. Musculoskeletal Translational Innovation Initiative, Department of Orthopedic Surgery, Beth Israel Deaconess Medical Center and Harvard Medical School, 330 Brookline Ave, RN123, Boston, MA, 02215 USA
  8. Harvard Medical School, Boston, MA 02115 USA
  9. Athinoula A Martinos Center for Biomed. Imaging, Massachusetts General Hospital 149 13, St. Charlestown, MA 02129 USA

Peer review process

Revised: This Reviewed Preprint has been revised by the authors in response to the previous round of peer review; the eLife assessment and the public reviews have been updated where necessary by the editors and peer reviewers.

Read more about eLife’s peer review process.

Editors

  • Reviewing Editor
    Peng Liu
    Shanxi Medical University, Taiyuan, China
  • Senior Editor
    Aleksandra Walczak
    École Normale Supérieure - PSL, Paris, France

Reviewer #1 (Public Review):

Summary:

In this work authors are trying to satisfy a real need in MR safety, when concerns can rise about the thermal increase due to metallic materials in patients carrying orthopedic implants. The "MR conditional" labeling of the implant obtained by ASTM in-vitro tests may help to plan the MR scan, but it is normally limited to a single specific MR sequence and a B0 value, and it is not always available. The adoption of an in-silico simulation testbed overcomes this limitation, providing a fast and reliable prediction of temperature increase from RF, in real-life scan conditions on human-like digital models. The FDA is pushing this approach.

Strengths:

The presented in-silico testbed looks valuable and validated. It is based on the widely available Visible Human Project (VHP) datasets, and the testbed is available on-line. The approval of the testbed by the FDA as a medical device development tool (MDDT) is a good premise for the large-scale adoption of this kind of solution.

Weaknesses:

A couple of limitations of the study are now clearly highlighted to the readers in this revised version of the paper. The following aspects:
- the lack of the equivalent modeling for the gradients-related heating;
- the way the implant is embedded in the VHP model that should take in consideration how to manage the removed and stretched tissues;
are now correctly taken in consideration in the discussion, providing additional literature.

Reviewer #2 (Public Review):

Summary:

In this article, the authors provide a method of evaluating safety of orthopedic implants in relation to Radiofrequency induced heating issues. The authors provide an open source computational heterogeneous human model and explain computational techniques in a finite element method solver to predict the RF induced temperature increase due to an orthopedic implant while being exposed to MRI RF fields at 1.5 T.

Strengths:

The open access computational human model along with their semiautomatic algorithm to position the implant can help realistically model the implant RF exposure in patient avoiding over- or under-estimation of RF heating measured using rectangular box phantoms such as ASTM phantom. Additionally, using numerical simulation to predict radiofrequency induced heating will be much easier compared to the experimental measurements in MRI scanner, especially when the scanner availability is limited.

Weaknesses:

The proposed method only used radiofrequency (RF) field exposure to evaluate the heating around the implant. However, in the case of bulky implants the rapidly changing gradient field can also produce significant heating due to large eddy currents. So the gradient induced heating still remains an issue to be evaluated to decide on the safety of the patient. Moreover, the method is limited to a single human model and might not be representative of patients with different age, sex and body weights.

Author Response

The following is the authors’ response to the original reviews.

The Authors wish to thank the Reviewers for their detailed and insightful comments. By properly addressing these critiques, we sincerely believe our finished product will be substantially improved and provide greater insight to the academic community.

Both Reviewers noted the importance of identifying the limitations of our study with particular emphasis on embedded implant heating due to switching gradient coils. Understanding the limitations of any model and/or simulation process is critical when adopting its use, especially when estimating the safety of embedded devices. For this reason, we have included the following text and corresponding references in our Discussion section:

While the workflow presented herein establishes a validated approach to estimate RF heating due to the presence of a passive implant within a human subject undergoing an MR procedure, certain limitations and proper use stipulations of this methodology should be identified. These include:

  1. The approach of embedding a given passive implant must be carefully considered and supervised by an orthopaedic subject matter expert, preferably an orthopaedic surgeon. While the procedures described above focus on insertion and registration of an implant to make it numerically suitable for simulation, relevant anatomic and physiological considerations must also be addressed to ensure a physically realistic and appropriate result. This will enable a proper simulated fit and no empty spaces or unintended tissue deformations.

  2. Temperature changes presented are due only to RF energy deposition. The results do not take into account the impact of low-frequency induction heating of metallic implants naturally caused by the switching gradient coils. Important work on this subject matter has recently been reported in [21],[22],[23],[24],[25],[26],[27]. Unless an orthopaedic implant has a loop path, heating due to gradient fields is typically less than heating due to RF energy deposition. The present testbed would be applicable to the induction heating of implants (and the expected temperature rise of nearby tissues), after switching from Ansys HFSS (the full wave electromagnetic FEM solver) to Ansys Maxwell (the eddy current FEM solver). Two examples of this kind have already been considered in [25],[45].

  3. The procedures presented in this work have been based on the response of a single human model of advanced age and high morbidity.

  4. Finally, validation was achieved using available published data [42]-[44] and relies upon the legitimacy and veracity of that data. Coil geometry, power settings, and other relevant parameters were taken explicitly from these sources and modeled to enable a faithful comparison.